Sterilizers for medical purposes - Low temperature vapourized hydrogen peroxide sterilizers - Requirements and testing
This document specifies requirements and tests for low temperature hydrogen peroxide sterilizers, using vaporized aqueous solution of hydrogen peroxide as the sterilizing agent. These sterilizers are used for the sterilization of medical devices, particularly thermolabile medical devices. This document specifies minimum requirements for: - the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices; - the equipment and controls of these sterilizers needed for operation, control, and monitoring, and which can be used for validation of the sterilization processes; - the test equipment and test procedures used to verify the sterilizer performance specified by this document. This document does not specify requirements for sterilizer intended to process liquids, biological waste, or human tissues unless part of a medical device. This document does not describe a quality management system for the control of all stages of the manufacture of the sterilizer.